Sample Throughput
1~8 samples at the same time
Certificate
CE; CB; RoHS
Pipetting Range
20μL ~250μL
Detection Time
1~2 hours,relying on the reagent
Channel and Available Fluorescein
Channel 1: FAM, SYBR Green I, etc. Channel 2: VIC, HEX, TET, JOE, etc.
Channel 3: ROX, Texas Red, etc. Channel 4: Cy5, etc.
Dimension
750mm(L) × 350mm (W) × 600mm (H)
Weight
80Kg
Pipetting Performance
20μL≤V<40μL: accuracy:A≤5.0%,repeatability:CV≤3.0%
40μL≤V<100μL: accuracy: A≤3.0%,repeatability: CV≤1.5%
V≥100μL: accuracy:A≤1.0%,repeatability:CV≤1.0%
Extraction Heating Rate
Average heating rate: ≥1.5℃/s
Extraction Temperature Accuracy
≤1.0℃
PCR Heating Rate
Average heating rate:≥ 4.5℃/s Maximum heating rate:≥ 6.1℃/s
PCR Cooling Rate
Average cooling rate: ≥ 3.5℃/s Maximum cooling rate:≥ 5.0℃/s
PCR Temperature Accuracy
≤0.1%
Sample Testing Linearity and Repeatabilit
Linear correlation: /r/≥ 0.998 Repeatability: CV ≤1.5%
Information Management
Sample information: with scanner inside Panall 8000 can scan and record the sample information
Reagent information: visual system can automatically identify the kit information and run the corresponding program
Minimized Contamination
Directional exhaust & negative pressure system
HEPA fltration UV disinfection
Shortest fixed path for sample operation
Data Storage
1000 experiment files can be stored
Language
Chinese and English
Power Supply and Power Consumption
AC 100~240V,50/60Hz:1200VA
Communication Specification
Internet Port: TCP/lP protocol, Ethernet connection ///USB Port: USB 2.0